3D-printed magnetic-based air pressure sensor for continuous respiration monitoring and breathing rehabilitation

Figure 3. Characterization of the fully assembled respiration sensor. (A) Output voltage (calibrated) plotted against increasing air pressure (in cmH2O); (B) Correlation plot between air pressure and magnet-to-sensor distance (experimental and calculated); (C) Calibrated output voltage for step-by-step loading and unloading of positive air pressure (inhalation); (D) Calibrated output voltage for step-by-step loading and unloading of negative air pressure (exhalation); (E) Output voltage for 1,000 cycles of continuous air pressure application alternating between 8 cmH2O and -15 cmH2O; (F) The first and final ten cycles of the output voltage plot in Figure 1E.