Method validation and comparison of quantification strategies for analysis of chlorinated paraffins in indoor dust by liquid chromatography and high-resolution mass spectrometry
Abstract
This paper describes the validation of a method for the simultaneous analysis of short-, medium-, and long-chained chlorinated paraffins (SCCPs, MCCPs, and LCCPs, respectively) in indoor dust by ultrasonic extraction and liquid chromatography quadrupole time-of-flight high-resolution mass spectrometry (LC-QTOF-HRMS). A series of spike and recovery experiments (n = 54) were conducted using CPs with varying carbon-chain lengths, chlorination degree, and concentrations. Technical standard mixtures of the SCCPs, MCCPs, and LCCPs were used to quantify spiking experiments by two commonly used calibration procedures: pattern deconvolution and chlorine-content calibration. The results quantified by pattern deconvolution meet the acceptability limits of the European Union Reference Laboratory (EURL) for all tests with trueness ranging from 72% to 141% and good precision represented by coefficients of variation (CVs) less than 15% in all experiments. The chlorine-content calibration also performed well overall, but on average overestimated concentrations for SCCPs and MCCPs by 32% and 25%, respectively, and did not meet the EURL’s trueness limits in all cases. CVs were below 18% for all results derived from the chlorine-content quantification. The final method was successfully applied to indoor dust samples from offices (n = 4), homes (n = 3), and a vehicle (n = 1) from Melbourne, Australia, with SCCPs (C10-13), MCCPs (C14-17), and
Keywords
INTRODUCTION
Chlorinated paraffins (CPs) are complex mixtures of polychlorinated n-alkanes with the general chemical formula CxH2x+2-yCly and whose chlorination degree can vary between 30% and 70% by weight (wt %). They are typically divided according to their carbon chain length into short chain CPs (SCCPs, C10-13), medium chain CPs (MCCPs, C14-17), and long chain CPs (LCCPs, C>17)[1]. CPs have been produced and used since the 1930s for a wide range of industrial applications including as additives in metalworking fluids and as flame retardants and plasticizers in polymers, paints, sealants and adhesives, textiles, and consumer products[2]. According to an investigation by Glüge et al.[3], CPs are produced in high volumes with an annual global production volume of more than 1 million tons since 2012, of which the worldwide production volume of SCCPs alone was estimated to be 165,000 tons per year.
CPs have been detected in a variety of environmental media[2,4-6] as well as human milk and serum[7,8]. Recent studies have shown that unintentional ingestion and inhalation of dust[5,6,9] and dietary intake[10,11] are among the most prominent pathways for human exposure to CPs. Due to their persistence, bioaccumulation potential, and toxic effects[12], international conventions and regulatory agencies have classified SCCPs as environmentally hazardous compounds and globally restricted their production[5,13]. The ban of SCCPs has thus caused a shift towards the production of MCCPs and LCCPs, whose safety is however also questioned[14,15].
CPs as applied within materials and products are composed of very complex mixtures of thousands of isomers differing by carbon chain length and chlorination degrees, which may also contain other hydrocarbon impurities[16]. Several analytical methods have been utilized, applying either gas or liquid chromatography (GC and LC, respectively) coupled with a range of mass spectrometric (MS) detectors. Among the most broadly applied methods has been GC-MS operated in single ion monitoring mode using electron capture negative ionization (ECNI)[16]. GC-ECNI/MS offers sensitive detection of [M-Cl]- and
HRMS systems such as quadrupole time-of-flight (QTOF) and Orbitrap-MS have also gained popularity for CP analysis due to their ability to provide congener group level separation of CPs via accurate mass identification not afforded by LRMS instruments[20,22,23]. While neither GC nor LC can fully separate congener groups chromatographically, robust data on CP patterns in samples may be generated by HRMS[18]. Atmospheric pressure chemical ionization (APCI) and electrospray ionization (ESI) modes have both been applied effectively for CP analysis by monitoring [M-H]-, [M+Cl]-, or other ions[10,24,25]. The post-column addition of dichloromethane to the mobile phase has been shown to favor the formation of [M+Cl]- ions to enhance sensitivity and selectivity[21,26], and the same effect was achieved more recently by
Another important challenge in achieving reliable CP quantification is a lack of commercially available, well-characterized standards suitable for analysis of each of the SCCP, MCCP, and LCCP groups. While several single compounds and single carbon-chain length standards are available, technical mixture standards are often used to cover the broad range of CP compositions possible in samples[28,29]. Standards for LCCPs are particularly limited. Several innovative quantification procedures have been proposed to allow quantification of CPs at the ∑SCCP, ∑MCCP, and ∑LCCP level, as well as chain-length and congener group levels[20]. Among the most commonly applied quantification strategies are pattern deconvolution, as first described by Bogdal et al.[21], and the chlorine-content calibration procedure introduced by Reth et al.[30], which both attempt to compensate for the variations in total CP response factors according to overall chlorine content. Pattern deconvolution achieves this by algorithmically matching homolog patterns in samples to those in a combination of CP mixture standards to determine appropriate response factors for quantification[21,25], while chlorine-content calibration utilizes the correlation between the measured chlorine content and the summed response of CP homologs to determine suitable response factors[30]. Specific challenges relating to the availability of suitable standards or particular behavior of CP responses are inherent to all proposed quantification procedures such that no one method has predominated[18].
The objective of this study was to validate a method for the analysis of SCCPs, MCCPs, and LCCPs in indoor dust by ultrasonic extraction and LC-QTOF-HRMS. Series of fortification experiments encompassing varying congener group compositions and total concentrations were quantified using two of the most commonly applied quantification procedures, namely pattern deconvolution and chlorine-content calibration, to assess the respective performance of these methods in terms of trueness and precision. Both calibration methods were applied to indoor dust from Melbourne, Australia, to assess the applicability of the overall procedure and illustrate the CP congener group profiles present in real samples.
EXPERIMENTAL
Standards and reagents
Technical mixture standards of SCCP 51.5%Cl (Lot: G976504CY), SCCP 55.5%Cl (Lot: G986911CY), SCCP 63%Cl (Lot: G986931CY), MCCP 42%Cl (Lot: 417125CY), MCCP 52%Cl (Lot: G162888CY), MCCP 57%Cl (Lot: G991196CY) LCCP (C18-20) 36%Cl (Lot: 114017CY), and LCCP (C18-20) 49%Cl (Lot: 855677CY) were purchased from Dr. Ehrenstorfer GmbH (Augsburg, Germany) and individual standards of
Sampling and sample extraction
Indoor dust was sampled from offices (n = 4), homes (n = 3), and a vehicle (n = 1), denoted as O, H, and V, respectively, in Melbourne, Australia during 2016, as described by McGrath et al.[31]. Briefly, nylon socks of pore size 25 μm were inserted into the inlet nozzle of a domestic vacuum cleaner and used to sample dust from 1 m2 of floor space in carpeted rooms and 4 m2 for rooms with hard floors. Dust was sampled with the vacuum cleaner from the front interior seats, doors, and dashboard surfaces of the vehicle. Samples were sieved to a < 500 μm fraction and stored at 4 °C in darkness prior to analysis.
Dust samples (50 mg) were weighed into 15 mL pre-cleaned glass vials, spiked with 20 ng of internal standard (IS) 13C-β-HBCD and extracted in 5 mL of a 3:1 mixture of n-hexane:dichloromethane using 1 min of vortexing and 10 min of ultrasonication. Extracts were centrifuged at 2000 rpm for 3 min, the supernatant transferred to clean vials, and the extraction repeated once more with fresh solvent. Extracts were purified by addition of 2 g of acidified silica (44% w/w), and then evaporated under nitrogen flow to incipient dryness and reconstituted in 0.5 mL n-hexane. The concentrated extracts were then loaded onto Agilent Bond Elut silica cartridges pre-conditioned by 6 mL of dichloromethane, positive pressure drying, and then 6 mL of n-hexane. Extracts were eluted from the cartridges using 6 mL of hexane (waste fraction) and 12 mL of dichloromethane (analysis fraction). The analysis fraction was evaporated under nitrogen and reconstituted to a final volume of 100 μL of methanol containing 20 ng of external standard (ES) d18-γ-HBCD. Procedural blanks (n = 4) and fortified procedural blanks (n = 54) were performed to the same specifications as dust samples without the presence of the dust matrix. Field blanks (n = 2) were collected with samples by “sampling” precleaned Na2SO4 from a piece of aluminum foil by the same method as dust collection. Field blanks were stored and analyzed in the same manner as dust samples.
Instrumental analysis
All analyses were performed on an Agilent 1290 Infinity II LC coupled to an Agilent 6560 QTOF-MS operated in electrospray ionization negative mode (ESI-). Injections of 5 μL were delivered onto an Agilent Zorbax Eclipse Plus C18 Rapid Resolution HD column (2.1 mm × 50 mm, 1.8 μm pore size) fitted with a matching guard column of 5 mm length. The mobile phase consisted of (A) Milli-Q water and (B) methanol containing 0.025 mM NH4Cl, which was applied using a gradient program that evolved linearly from 10% B to 95% B from 0 to 3 min, held at 95% B from 3 to 9 min, returned linearly to 10% B from 9 to 10 min, and remained at 10% B for 2 min. The flow rate was 0.2 mL/min, and the column temperature was maintained at 40 °C. MS acquisition was performed in full scan mode from 100 to 1700 m/z using real-time calibration of mass accuracy via reference standards of purine (119.0360 m/z) and hexakis phosphazine (966.0007 m/z) delivered to the MS during analysis. The MS gas temperature was 300 °C, the drying gas flow was 8 L/min, the nebulizer pressure was 40 psi, the sheath gas temperature was 300 °C, and the sheath gas flow rate was 6.5 L/min. The fragmentor and nozzle voltages were set to 300 and 100 V, respectively.
Analytes were identified and integrated using Agilent MassHunter Quantitative Analysis Version 10.0. The first and second most abundant isotopes for [M+Cl]- ions representing CP homologs ranging from C10 to C25 and Cl4 to Cl12 were extracted with a mass accuracy tolerance of 20 ppm [Supplementary Table 1 and 2]. Analytes were considered detected when the retention time of peaks matched those of technical mixture standards within ±5%, isotope ratios were within ±20% of theoretical values, and the signal to noise ratio exceeded 3.
Quantification
Two separate quantification strategies were applied to the acquired sample data to assess the applicability of the methods. The pattern deconvolution quantification procedure was first described by Bogdal et al.[21] and was applied in this study using the open-source programming code provided by Perkons et al.[24] in RStudio software Version 1.2.1335. Six-point calibrations at total concentrations ranging from 0.125 to 4.0 μg/mL were prepared separately for each of the eight CP technical standard mixtures, each containing the IS and ES at 200 ng/mL. Relative response factors were determined for each of the analyzed CP homolog groups from the eight CP calibration curves according to the total CP concentration in each calibrant. The pattern deconvolution algorithm provided by Perkons et al.[24] was then used to determine the linear combination of the prepared reference technical mixture standards that best approximated the homolog relative response distribution measured in samples. ∑SCCP, ∑MCCP, and ∑LCCP quantification was then performed using the supplied algorithm to apply a combination of relative response factors which corresponds to the proportions of individual CP technical mixtures utilized for the pattern reconstruction. A detailed description of the pattern deconvolution procedure was described by Perkons et al.[24], and the R code applied is available in GitHub (https://github.com/ingusperkons/CP-Crawler).
The chlorine-content calibration procedure was modified from the methods first described by Reth et al.[30] for quantification of SCCPs and MCCPs from GC-ECNI/MS analysis data. In the present study, technical CP standards were used to prepare calibration curves at five degrees of chlorination separately for each of the SCCP, MCCP, and LCCP groups at fixed ∑SCCP, ∑MCCP, and ∑LCCP concentrations of 4.0 μg/mL each, respectively. Each calibrant contained the IS and ES at 200 ng/mL to correspond with concentrations added to samples. Relative response factors were derived from the summed relative responses of CP homologs measured in calibrants, plotted against calculated chlorine-contents in respective standards, and fitted with exponential regression curves. The chlorine-contents and total relative response measured in samples were then used to derive ∑SCCP, ∑MCCP, and ∑LCCP concentrations using the respective calibration equations. Full details of the chlorine-content calibration procedure are given by
Response of the 13C-β-HBCD IS was used to derive relative responses for individual CP homolog groups for both applied quantification procedures. While suitable isotopically labeled CP standards remain commercially unavailable[28], 13C-β-HBCD was selected for use as an IS as it behaves similarly to CPs during sample preparation (e.g., it is found in the same fraction as the CPs during the silica fractionation step) and has been found to perform well in correcting for extraction losses and analytical variability[10].
Statistics
Statistical calculations were performed using Microsoft Excel 16. Percentage trueness was defined in fortified blank tests as the measured concentration in extracts divided by the known fortification concentration multiplied by 100. Precision was expressed as the coefficient of variation (CV) between measurements in triplicated fortified blank tests, i.e., the standard deviation between measurements divided by the average of measurements multiplied by 100. For assessment of the difference in values in dust samples as calculated by pattern deconvolution and chlorine-content calibration procedures, percentage differences were determined as the absolute difference between the two values divided by the average of the values multiplied by 100.
Estimates of CP intake via ingestion of indoor dust were calculated by the procedure described by
RESULTS AND DISCUSSION
Technical standard impurities and implications for method validation
A common and practical way to assess the accuracy and precision of an analytical procedure is to fortify a representative sample matrix with known levels of a given analyte and compare the measurement with the known, spiked concentration. Assessment of methods for CP analysis is, however, complicated by the lack of availability of well-defined standards. The individual technical standards used for calibration in this study were found to contain homologs not consistent with the defined SCCP, MCCP, and LCCP carbon-chain groupings. For example, on the basis of summed relative response, SCCPs only accounted for 71% of the CPs measured in the SCCP 55.5%Cl standard, with MCCPs and LCCPs constituting 17% and 12% of the mixture, respectively. The MCCP 52%Cl standard contained traces of SCCPs and LCCPs, each < 1%, while the LCCP 49%Cl standard mixture was found to contain the MCCP C17 homolog at a proportion of 29% of summed relative response. Furthermore, despite the LCCP standards having stated carbon-chain length compositions of only C18-20 homologs, congener groups of C21 and C22 were determined at low levels of each < 1% in the LCCP 49%Cl standard and homologs up to C25 were present in the LCCP 36%Cl standard.
Similar findings in technical standards from the same manufacturer, Dr. Ehrenstorfer, have been evidenced in other studies[27,32,33]. Zheng et al.[27] reported proportions of MCCPs at rates of up to 10.5% in SCCP technical mixtures and ranging from 11.2% to 29.3% in the LCCP standards. Separate analyses by Li et al.[33] and Huang et al.[32] each determined substantial levels of the MCCP C17 homologs in LCCP mixtures at proportions of 7%-21% and 13.5%-30.6%, respectively, while C21-27 homologs were also detected in the
Since the exact proportion of CP impurities in the technical standards cannot be determined, spike and recovery tests must necessarily be performed from separate spikes of the SCCP, MCCP, and LCCP groups so the fortification level of any individual group can be known. In the present study, validation experiments were designed to assess the trueness and precision of the analytical method quantified by both pattern deconvolution and chlorine-content calibration procedures in a series of fortified procedural blank extractions. Due to the ubiquitously high levels of CPs measured in indoor dust, a “clean” dust matrix with which to conduct fortification testing was not available[34]. Nine individual CP spike mixtures, denoted A-I, were prepared to represent three different chlorine compositions for each of the SCCP, MCCP, and LCCP homolog groups, as detailed in Supplementary Table 3. Fortified blank extraction tests were performed for each of the A-I mixtures in triplicate at two concentration levels to total 54 individual extraction and analysis experiments. Low spike tests contained 100 ng (equivalent to a concentration of 2.0 μg/g dust based on a 50 mg dust extraction) and a high spike of 400 ng (equivalent to 8.0 μg/g dust) of ∑SCCPs, ∑MCCPs, or ∑LCCPs, respectively.
Trueness and precision
The results of the fortified blank assessment are shown in Figure 1. For SCCPs, individual measurements as quantified by the pattern deconvolution procedure ranged from 72% to 117% trueness among tests at both high and low spike levels. Trueness values were each within the acceptability limits of 50%-150% as applied in recent proficiency testing of CP analysis in food by the European Union Reference Laboratory (EURL)[35,36]. Average SCCP measurements determined by the chlorine-content procedure were higher than those of the pattern deconvolution method for each of the fortification scenarios with trueness among the individual samples ranging from 93% to 174% (4 of 18 measurements outside the EURL criteria). High precision was observed in the results obtained by both quantification procedures for SCCPs with CVs ≤ 15% in all cases. Similar results were observed for the MCCP assessment, with trueness derived from pattern deconvolution again meeting the EURL criteria in all instances with a range of 85%-141% among individual measurements. MCCP values quantified by the chlorine-content quantification typically overestimated concentrations in the spike analyses, with trueness ranging 85%-178% and all but the low-level E spikes meeting the EURL criteria. Comparisons of MCCP CV values for each of the scenarios demonstrated marginally better precision in the pattern deconvolution results, all ≤ 13%, while CVs in the tests as computed by chlorine-content calibration were also acceptable with a maximum of 18%. The results for LCCP tests considering only the C18-20 homolog groups stated within standard concentrations are remarkably similar for both quantification strategies, with individual trueness ranging 80%-129% and 77%-124% for the pattern deconvolution and chlorine-content outputs, respectively. Precision was also similar between the two procedures and markedly better than that seen in the SCCP and MCCP analyses with the majority of CV values ≤ 8% and ≤ 13%, respectively.
Figure 1. Trueness (%) of ∑SCCP, ∑MCCP, and ∑LCCP measurements of fortified blank extraction as quantified by pattern deconvolution or chlorine-content calibration. Low and high spike levels were 100 and 400 ng, respectively, of ∑SCCPs, ∑MCCPs, or
Both the pattern deconvolution and chlorine-content approaches have been applied successfully for the analysis CPs in a wide range of matrices, although previous validation of these methods has not typically compared the calibration capabilities directly[17,23,24,37]. A range of 91%-123% trueness was recorded by
Limits of detection and quantification
In this study, limits of detection (LODs) were determined as the lowest concentration of ∑SCCP, ∑MCCP, and ∑LCCP in which at least one congener group can be detected in both the lowest and highest Cl% technical standard mixtures. SCCP, MCCP, and LCCP LODs determined by these criteria were 10, 1.0, and 5.0 ng/mL, respectively, in calibrant solutions, corresponding to dust concentrations of 20, 2.0, and 10 ng/g [Table 1]. With regards to instrument sensitivity, these LODs are similar to those reported for other chlorine-enhanced LC-ESI-QTOF methods[27,39] and slightly higher than those determined for SCCP and MCCP analysis by GC/ECNI-Orbitrap-MS[40].
Limits of detection and quantification (LODs and LOQs, respectively) for SCCPs, MCCPs, and LCCPs
Instrument (ng/mL) | Dust (ng/g) | |||
LOD | LOQ | LOD | LOQ | |
SCCP | 10 | 100 | 20 | 200 |
MCCP | 1.0 | 50 | 2.0 | 100 |
LCCP | 5.0 | 50 | 10 | 100 |
Limits of quantification (LOQs) were defined as the lowest ∑SCCP, ∑MCCP, and ∑LCCPs concentration at which at least half of the congeners detected in the highest concentration calibrants were also detected in both the lowest and highest Cl% standard mixtures. These criteria were selected following technical mixture standard dilution series analysis that revealed this to be the minimum level of congener group identification required for effective chlorine-content determination. Instrumental LOQs for SCCPs, MCCPs, and LCCPs were 100, 50, and 50 ng/mL, respectively, corresponding to concentrations of 200, 100, and 100 ng/g in dust, which were, again, similar to those of previous LC-ESI-QTOF analyses[27]. Lower LOQs were reported for SCCPs and MCCPs using GC-ECNI-Orbitrap[40] and SCCPs and LCCPs using LC-APCI-HRMS[1], but marginally higher ones were reported for GC-ECNI-LRMS[17].
Procedural blanks extracted and analyzed with each batch of 20 samples contained low but consistent levels of some C14 congener groups of Cl5-7 and C15 of Cl6-7, while SCCPs and LCCPs were not detected. The number of congener groups detected in procedural blanks was below the proportion required for reliable quantification, and relative responses were substantially lower than those in the described MCCP LOQs, thus no blank subtractions were deemed necessary. Analysis of field blanks revealed that the sampling protocol did not contribute CP contamination to the analyses. The IS was well recovered from all procedural blanks, fortified procedural blanks, and dust samples with an average (± CV) of 101% ± 16%. Acceptable limits of IS recovery for the method were defined as 50%-130%, such that recovery was not factored into the LOQ determinations.
Analysis of real samples
The applicability of the analytical method was demonstrated via simultaneous measurements of SCCPs, MCCPs, and LCCPs in eight real dust samples from offices, homes, and a vehicle from Melbourne, Australia. Concentrations derived by the pattern deconvolution and chlorine-content quantification methods were similar, as shown in Figure 2. ∑SCCPs ranged from 7.7 to 100 μg/g (median, 38 μg/g) by pattern deconvolution and 4.6 to 130 μg/g (median, 55 μg/g) by chlorine-content calibration. ∑MCCP concentrations were the dominant homolog group in all samples ranging from 62 to 240 μg/g (median,
Figure 2. Concentrations of ∑SCCPs, ∑MCCPs, and ∑LCCPs determined by pattern deconvolution (PD) and chlorine-content calibration (CC) in office (O), home (H), and vehicle (V) dust samples from Melbourne, Australia.
Overall, the percentage differences in ∑CP concentrations in dust between the two quantification procedures had a maximum of 22% (O-3), while most samples differed by ≤ 15%, and two samples varied only by 1% each (O-4 and V-1). No particular trends could be discerned for levels of SCCPs, MCCPs, or LCCPs among the three indoor environment types (office, home, and vehicle).
The levels of CPs measured in the current study spanned a similar range to those determined in 44 samples of indoor dust from houses, offices, and vehicles in Brisbane and Canberra, Australia[6]. MCCPs were also the dominant group in the analysis by He et al.[6], ranging from 5.1 to 540 μg/g with a median of 95 μg/g. SCCPs ranged from 0.29 to 58 μg/g and LCCPs were detected in 86% of samples up to a maximum concentration of 27 μg/g in the same study, which reported overall concentrations to be generally greatest in vehicles, followed by offices and then houses. Studies of indoor dust from Canada[9], Norway[1], South Africa[5,41], and China have each demonstrated similar predominance of MCCPs and ∑CP concentrations in the mid-microgram per gram range, suggesting that CPs are some of the most prevalent organohalogen contaminants in the indoor environment[34].
The carbon-chain and chlorine number distribution of congener groups based on relative response in all dust samples are shown in Figure 3, with full congener group patterns shown for selected samples in Figure 4. While appropriate analytical standards are not available to allow for congener group-specific quantification, a comparison of the relative responses among the groups provides an indication as to the distribution of CP homologs, although the difference in compound sensitivity cannot be accounted for in these comparisons. SCCPs were dominated by the ∑C13 homologs, which accounted for approximately half of total relative response, with smaller levels of ∑C12 and ∑C11 and lower levels again of ∑C10. The distribution of carbon-chain groups was more even for MCCPs with levels of ∑C14 marginally higher than other groups, accounting for around 30%-45% of total relative response and the C15-17 congener groups present at similar levels. Of the LCCPs with stated concentration in the standards used for quantification, the ∑C18, ∑C19, and ∑C20 relative response levels were approximately equal, while these congener groups alone only accounted for approximately 30%-60% of the total relative response among the C18-25 LCCPs measured. All C21-25 LCCP carbon-chain groupings were detected in each of the eight dust samples. While proportions differed by sample, summed relative responses for the ∑C21 and ∑C22 homologs were each only slightly lower than those of the ∑C18-20 groups, with the ∑C23, ∑C24, and ∑C25 groups slightly lower again. In at least one sample, O-1, the C21-25 groups not incorporated into ∑LCCP quantification dominated the overall relative response [Figure 4]. The distribution of chlorine number was mostly consistent between the dust samples for SCCPs and MCCPs, with a dominance of Cl7-8 observed in both groups. Greater variation in Cl distribution for LCCPs between dust samples was evident, but with Cl7-10 groups dominant overall.
Figure 3. Proportional distribution of carbon-chain groups (A) and chlorine number (B) in office (O), home (H), and vehicle (V) dust from Melbourne, Australia, by relative response of individual CP congener groups.
Figure 4. Relative responses of individual CP congener groups in selected office (O), home (H), and vehicle (V) indoor dust samples from Melbourne, Australia.
The carbon-chain and chlorine distributions of CPs in dust have typically mirrored patterns observed in known commercial formulations of CPs[41] but may also be influenced by environmental processes affecting release rates from indoor sources and partitioning between dust and air[1]. The dominance of C13 homologs among SCCPs and C14 among the MCCPs measured in the Melbourne dust samples appears to correspond to a prevalence of these groups in CP-42 and CP-52 mixtures[42]. Various LCCP homolog patterns of C18-30 were observed among the three CP-52 and three CP-70 technical products analyzed by
Human exposure assessment
The concentrations of CPs measured in dust samples in the present study were used to calculate estimates of CP intake via indoor dust for adults and toddlers in Melbourne, Australia [Table 2]. Given the small number of samples, intake estimates were only computed at the median concentration level and provide a preliminary indication of exposure via dust which excludes contributions from time spent in vehicles. ∑CP(C10-20) intake by adults according to mean and high dust ingestion scenarios was estimated to be 40 and 100 ng/kg bw/day, respectively, while toddlers were expected to experience higher exposure rates of 410 and 1700 ng/kg bw/day for the mean and high dust ingestion scenarios, respectively. Rates of exposure to
Estimates of CP intake (ng/kg bw/day) via indoor dust ingestion for adults and toddlers
Age group | Dust ingestion scenario | ∑SCCPs | ∑MCCPs | ∑LCCPs (C18-20) | ∑CPs (C10-20) |
Adult | Mean | 6.1 | 30 | 6.1 | 40 |
High | 15 | 76 | 15 | 100 | |
Toddler | Mean | 63 | 350 | 56 | 410 |
High | 250 | 1400 | 220 | 1700 |
CONCLUSIONS
This study provides a comprehensive method validation for the extraction and analysis of SCCPs, MCCPs, and LCCPs in indoor dust by LC-QTOF-HRMS. A direct comparison of two common quantification strategies for CPs was performed by applying each of the protocols to the same set of LC-QTOF acquisition data so that the relative strengths of pattern deconvolution and chlorine-content calibration could be evaluated. This assessment showed that results generated by pattern deconvolution quantification met the EURL’s accuracy criteria in all applied tests and exceeded the performance of the chlorine-content procedure in terms of both trueness and precision measures. Application of the analytical method to real indoor dust samples demonstrated the sensitive and simultaneous measurement of SCCPs, MCCPs, and LCCPs with low levels of deviation between results obtained by the two quantification methods. This first report of CP concentrations in indoor dust from Melbourne, Australia, found MCCPs to be dominant, while ubiquitous detection of LCCP congener groups extending to carbon-chain lengths outside the range of all available commercial standards exposed the strong need for appropriate analytical standards. A preliminary human exposure assessment indicated that CP intake via ingestion of indoor dust in Melbourne might represent a significant exposure pathway for both adults and toddlers.
DECLARATIONS
Authors’ contributions
Conceptualization, methodology, formal analysis, investigation, writing - original draft preparation, review and editing, visualization, funding acquisition: McGrath TJ
Conceptualization, methodology, resources, writing - review and editing, supervision, project administration, funding acquisition: Covaci A
Conceptualization, writing - original draft, review and editing, visualization, funding acquisition: Poma G
Availability of data and materials
Additional data for this study are presented in the supplementary information.
Financial support and sponsorship
The financial support was provided from a post-doctoral fellowship for Thomas J. McGrath from the Research Scientific Foundation-Flanders (FWO) - project number 12Z9320N and the University of Antwerp.
Conflicts of interest
All authors declared that there are no conflicts of interest.
Ethical approval and consent to participate
Not applicable.
Consent for Publication
Not applicable.
Copyright
© The Author(s) 2022.
Supplementary Materials
REFERENCES
1. Yuan B, Tay JH, Padilla-Sánchez JA, Papadopoulou E, Haug LS, de Wit CA. Human exposure to chlorinated paraffins via inhalation and dust ingestion in a Norwegian Cohort. Environ Sci Technol 2021;55:1145-54.
2. van Mourik LM, Gaus C, Leonards PEG, de Boer J. Chlorinated paraffins in the environment: a review on their production, fate, levels and trends between 2010 and 2015. Chemosphere 2016;155:415-28.
3. Glüge J, Wang Z, Bogdal C, Scheringer M, Hungerbühler K. Global production, use, and emission volumes of short-chain chlorinated paraffins - a minimum scenario. Sci Total Environ 2016;573:1132-46.
4. Gao W, Cao D, Wang Y, et al. External exposure to short- and medium-chain chlorinated paraffins for the general population in Beijing, China. Environ Sci Technol 2018;52:32-9.
5. Brits M, de Boer J, Rohwer ER, De Vos J, Weiss JM, Brandsma SH. Short-, medium-, and long-chain chlorinated paraffins in South African indoor dust and cat hair. Chemosphere 2020;238:124643.
6. He C, Brandsma SH, Jiang H, et al. Chlorinated paraffins in indoor dust from Australia: levels, congener patterns and preliminary assessment of human exposure. Sci Total Environ 2019;682:318-23.
7. Krätschmer K, Malisch R, Vetter W. Chlorinated paraffin levels in relation to other persistent organic pollutants found in pooled human milk samples from primiparous mothers in 53 countries. Environ Health Perspect 2021;129:87004.
8. Zhao N, Fang X, Zhang S, Zhu Y, Ding L, Xu C. Male renal functions are associated with serum short- and medium-chain chlorinated paraffins in residents from Jinan, China. Environ Int 2021;153:106514.
9. Shang H, Fan X, Kubwabo C, Rasmussen PE. Short-chain and medium-chain chlorinated paraffins in Canadian house dust and NIST SRM 2585. Environ Sci Pollut Res Int 2019;26:7453-62.
10. McGrath TJ, Limonier F, Poma G, et al. Concentrations and distribution of chlorinated paraffins in Belgian foods. Environ Pollut 2021;291:118236.
11. Krätschmer K, Schächtele A, Vetter W. Short- and medium-chain chlorinated paraffin exposure in South Germany: a total diet, meal and market basket study. Environ Pollut 2021;272:116019.
12. Wang X, Zhu J, Xue Z, Jin X, Jin Y, Fu Z. The environmental distribution and toxicity of short-chain chlorinated paraffins and underlying mechanisms: Implications for further toxicological investigation. Sci Total Environ 2019;695:133834.
13. SC-8/11: Listing of short-chain chlorinated paraffins. Available from: http://chm.pops.int/Portals/0/download.aspx?d=UNEP-POPS-COP.8-SC-8-11.English.pdf [Last accessed on 14 Mar 2022].
14. Guida Y, Capella R, Weber R. Chlorinated paraffins in the technosphere: a review of available information and data gaps demonstrating the need to support the Stockholm Convention implementation. Emerging Contaminants 2020;6:143-54.
15. Vetter W, Sprengel J, Krätschmer K. Chlorinated paraffins - a historical consideration including remarks on their complexity. Chemosphere 2022;287:132032.
16. van Mourik LM, Leonards PE, Gaus C, de Boer J. Recent developments in capabilities for analysing chlorinated paraffins in environmental matrices: a review. Chemosphere 2015;136:259-72.
17. Mcgrath TJ, Poma G, Bombeke J, et al. Optimization and validation of an analytical method for the quantification of short- and medium-chained chlorinated paraffins in food by gas chromatography-mass spectrometry. Food Control 2021;119:107463.
18. Yuan B, Muir D, MacLeod M. Methods for trace analysis of short-, medium-, and long-chain chlorinated paraffins: Critical review and recommendations. Anal Chim Acta 2019;1074:16-32.
19. Zencak Z, Oehme M. Recent developments in the analysis of chlorinated paraffins. Trends Analyt Chem 2006;25:310-7.
20. Mézière M, Krätschmer K, Pe Rkons I, et al. Addressing main challenges regarding short- and medium-chain chlorinated paraffin analysis using GC/ECNI-MS and LC/ESI-MS methods. J Am Soc Mass Spectrom 2020;31:1885-95.
21. Bogdal C, Alsberg T, Diefenbacher PS, MacLeod M, Berger U. Fast quantification of chlorinated paraffins in environmental samples by direct injection high-resolution mass spectrometry with pattern deconvolution. Anal Chem 2015;87:2852-60.
22. Mézière M, Cariou R, Larvor F, et al. Optimized characterization of short-, medium, and long-chain chlorinated paraffins in liquid chromatography-high resolution mass spectrometry. J Chromatogr A 2020;1619:460927.
23. van Mourik LM, Lava R, O’Brien J, Leonards PEG, de Boer J, Ricci M. The underlying challenges that arise when analysing short-chain chlorinated paraffins in environmental matrices. J Chromatogr A 2020;1610:460550.
24. Perkons I, Pasecnaja E, Zacs D. The impact of baking on chlorinated paraffins: Characterization of C10-C17 chlorinated paraffins in oven-baked pastry products and unprocessed pastry dough by HPLC-ESI-Q-TOF-MS. Food Chem 2019;298:125100.
25. Yuan B, Alsberg T, Bogdal C, et al. Deconvolution of soft ionization mass spectra of chlorinated paraffins to resolve congener groups. Anal Chem 2016;88:8980-8.
26. Zencak Z, Oehme M. Chloride-enhanced atmospheric pressure chemical ionization mass spectrometry of polychlorinated n-alkanes. Rapid Commun Mass Spectrom 2004;18:2235-40.
27. Zheng L, Lian L, Nie J, et al. Development of an ammonium chloride-enhanced thermal-assisted-ESI LC-HRMS method for the characterization of chlorinated paraffins. Environ Pollut 2019;255:113303.
28. Fernandes AR, Vetter W, Dirks C, et al. Determination of chlorinated paraffins (CPs): Analytical conundrums and the pressing need for reliable and relevant standards. Chemosphere 2022;286:131878.
29. Schinkel L, Bogdal C, Canonica E, et al. Analysis of medium-chain and long-chain chlorinated paraffins: the urgent need for more specific analytical standards. Environ Sci Technol Lett 2018;5:708-17.
30. Reth M, Zencak Z, Oehme M. New quantification procedure for the analysis of chlorinated paraffins using electron capture negative ionization mass spectrometry. J Chromatogr A 2005;1081:225-31.
31. McGrath TJ, Morrison PD, Ball AS, Clarke BO. Concentrations of legacy and novel brominated flame retardants in indoor dust in Melbourne, Australia: an assessment of human exposure. Environ Int 2018;113:191-201.
32. Huang X, Ding C, Su Q, et al. A simplified method for determination of short-, medium-, and long-chain chlorinated paraffins using tetramethyl ammonium chloride as mobile phase modifier. J Chromatogr A 2021;1642:462002.
33. Li T, Wan Y, Gao S, Wang B, Hu J. High-throughput determination and characterization of short-, medium-, and long-chain chlorinated paraffins in human blood. Environ Sci Technol 2017;51:3346-54.
34. Poma G, Mcgrath TJ, Christia C, Malarvannan G, Covaci A. Emerging halogenated flame retardants in the indoor environment. Emerging Halogenated Flame Retardants in the Environment. Elsevier; 2020. p. 107-40.
35. General Protocol for EU Proficiency Tests on Pesticide Residues in Food and Feed. Available from: http://www.eurl-pesticides.eu/userfiles/file//General_Protocol_v8.pdf [Last accessed on 14 Mar 2022].
36. Proficiency Test on the Determination of PCDD/Fs, PCBs, BFRs, PFASs and CPs in Fish fillet 2020. Available from: https://eurl-pops.eu/user/pages/06.news/23.EURL-PT-POP-2001-FI-PBDE/EURL-PT_Fish_fillet_2020-Report%20PBDE%20HBCDD%20v1.0.pdf?g-73eaced6 [Last accessed on 14 Mar 2022].
37. Brandsma SH, van Mourik L, O’Brien JW, et al. Medium-chain chlorinated paraffins (CPs) dominate in Australian sewage sludge. Environ Sci Technol 2017;51:3364-72.
38. Krätschmer K, Schächtele A. Interlaboratory studies on chlorinated paraffins: Evaluation of different methods for food matrices. Chemosphere 2019;234:252-9.
39. Zhou X, Wu H, Huang X, Hang F, Luo H. Development a simple and rapid HPLC-ESI-Q-TOF/MS method for determination of short- and medium- chain chlorinated paraffins in human serum. J Chromatogr B Analyt Technol Biomed Life Sci 2019;1126-1127:121722.
40. Krätschmer K, Cojocariu C, Schächtele A, Malisch R, Vetter W. Chlorinated paraffin analysis by gas chromatography Orbitrap high-resolution mass spectrometry: Method performance, investigation of possible interferences and analysis of fish samples. J Chromatogr A 2018;1539:53-61.
41. Chen H, Lam JCW, Zhu M, et al. Combined effects of dust and dietary exposure of occupational workers and local residents to short- and medium-chain chlorinated paraffins in a mega E-waste recycling industrial park in South China. Environ Sci Technol 2018;52:11510-9.
42. Li T, Gao S, Ben Y, Zhang H, Kang Q, Wan Y. Screening of chlorinated paraffins and unsaturated analogues in commercial mixtures: confirmation of their occurrences in the atmosphere. Environ Sci Technol 2018;52:1862-70.
43. Fridén UE, McLachlan MS, Berger U. Chlorinated paraffins in indoor air and dust: concentrations, congener patterns, and human exposure. Environ Int 2011;37:1169-74.
44. Liu LH, Ma WL, Liu LY, et al. Occurrence, sources and human exposure assessment of SCCPs in indoor dust of northeast China. Environ Pollut 2017;225:232-43.
Cite This Article
How to Cite
McGrath, T. J.; Covaci, A.; Poma, G. Method validation and comparison of quantification strategies for analysis of chlorinated paraffins in indoor dust by liquid chromatography and high-resolution mass spectrometry. J. Environ. Expo. Assess. 2022, 1, 9. http://dx.doi.org/10.20517/jeea.2021.10
Download Citation
Export Citation File:
Type of Import
Tips on Downloading Citation
Citation Manager File Format
Type of Import
Direct Import: When the Direct Import option is selected (the default state), a dialogue box will give you the option to Save or Open the downloaded citation data. Choosing Open will either launch your citation manager or give you a choice of applications with which to use the metadata. The Save option saves the file locally for later use.
Indirect Import: When the Indirect Import option is selected, the metadata is displayed and may be copied and pasted as needed.
Comments
Comments must be written in English. Spam, offensive content, impersonation, and private information will not be permitted. If any comment is reported and identified as inappropriate content by OAE staff, the comment will be removed without notice. If you have any queries or need any help, please contact us at support@oaepublish.com.